Honeywell UDA2182 Manuale Utente Pagina 157

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 258
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 156
Input Calibration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 145
Step Action Screen
12
Once the reading is stable
Press
Enter
“Buffer 2 stability check”
Use
to change the value of
the Buffer.
“Up/Down changes Buffer”
13
If the calibration fails, an error
message will be displayed
across the bottom stripe of the
screen.
Make necessary adjustments
and re-calibrate.
Error Messages:
Buffer span too low
OFFSET UNDERRANGE
OFFSET OVERRANGE
Slope underrange
Slope overrange
Solution Unstable
Temp Too Low
Temp too High
See Table 12-2 for definitions
8.5.3 Buffering Method of Calibrating pH Electrodes
Recommended for most applications
This technique is recommended for best accuracy in most applications.
Materials
Materials required are:
Two standard buffer reference solutions that are at least 2 pH apart from one another.
A container for each, large enough to immerse the electrode to measuring depth.
Distilled or de-ionized water to rinse the electrode.
Procedure
Make sure you have selected “PV Type –pH Glass, pH Durafet, or pH HPW ” in the
Inputs configuration - Table 6-5.
Refer to Section 6.4.1 –
General Rules for Editing.
Vedere la pagina 156
1 2 ... 152 153 154 155 156 157 158 159 160 161 162 ... 257 258

Commenti su questo manuale

Nessun commento