Temperature Input Calibration
188 UDA2182 Universal Dual Analyzer Product Manual January 2009
Viewing and resetting Temperature Offset
If the calibration is suspect, you can reset the Temperature Offset and calibrate again.
In the same screen as “
Temp Cal”, use the keys to highlight “Reset Tmp Offset”.
T1 pH/ORP CAL
Temp Cal
Temp Offset
(Read only)
Reset Tmp Offset
T1 pH/ORP CAL
Temp Cal
Temp Offset
(Read only)
Reset Tmp Offset
Figure 10-1 Resetting temperature offset
Press ENTER. The Temperature Offset will be reset to (default).