Calibration History
190 UDA2182 Universal Dual Analyzer Product Manual January 2009
11.2 Clear Calibration History
Press
Setup
to display the Main menu.
Use the keys to select “Maintenance” then press
Enter
to enter the sub-
menu.
Use the keys to select “Display” then press
Enter
to enter the sub-menu.
Use the keys to select “Clr Cal Hist” then press
Enter
to allow change.
Use the keys to select “Yes” then press
Enter
to clear the Calibration History
screen.