Input Calibration
162 UDA2182 Universal Dual Analyzer Product Manual January 2009
8.7.6 Resetting Calibration Trim
If the calibration is suspect, you can reset the Calibration Trim and calibrate again.
In the same screen as “
Sample Cal”, use the keys to highlight “Reset Trim”.
IN2 Conduc Cal
Sample Cal
Cal Trim
Reset Cal Trim
1.00
IN2 Conduc Cal
Sample Cal
Cal Trim
Reset Cal Trim
1.00
Figure 8-3 Resetting Calibration Trim
Press ENTER. The Calibration Trim will be reset to 1.00 (default).