Outputs Calibration
182 UDA2182 Universal Dual Analyzer Product Manual January 2009
Step Action Screen
8
Press
Enter
OUTPUT CAL
Output 1
Output 2
Output 3
OUTPUT CAL
Output 1
Output 2
Output 3
Use
to select
an Analog Output to be
calibrated
9
Press
Enter
OUTPUT 1
20mA Offset
4mA Offset
0
0
Reset 20mA Offs
Reset 4mA Offs
OUTPUT 1
20mA Offset
4mA Offset
0
0
Reset 20mA Offs
Reset 4mA Offs
Use
to select
” 20 mA Offset”
10
Press
Enter
OUTPUT 1
20mA Offset
4mA Offset
-147
3
Reset 20mA Offs
Reset 4mA Offs
OUTPUT 1
20mA Offset
4mA Offset
-147
3
Reset 20mA Offs
Reset 4mA Offs
The right most digit will be
“blinking”.
11
To correct the value on the
meter:
Use the keys to
increment or decrement the
value of the digit
Use the keys to move
the cursor to the next digit.
Repeat as required to
achieve a 20mA reading on
the test meter
When all digits have been
changed, press “Enter” to
store the 20mA value.
Press “Exit” to cancel. The
previous value is retained.
Vedere la pagina 193